Receptors
|
Silicon photocells
|
Measurement area
|
Ø27mm (Ø10mm probes also available)
|
Ø27mm
|
Acceptance angle
|
±2.5°
|
±2.5°
|
Measurement distance
|
30±10mm
|
30±10mm
|
Display range
|
Luminance
|
0.0001 to 1000 cd/m²
|
0.0001 to 1000 cd/m²
|
Chromaticity
|
Displayed as 4- or 3-digit value (can be chosen)
|
Luminance
|
Measurement range
|
0.0050 to 1000 cd/m²
|
0.0050 to 1000 cd/m²
|
Accuracy (white) *1
(temperature: 23° +-2°,
relative humidity
(40+-10_%))
|
0.0050 to 0.0999 cd/m2 (±4%±0.0015 cd/m2)
0.1000 to 9.999 cd/m2 (±3%±0.0010 cd/m2)
10.00 to 1000 cd/m2 (±2%±0.0010 cd/m2)
|
0.0050 to 0.0999 cd/m2 (±4%±0.0015 cd/m2)
0.1000 to 9.999 cd/m2 (±3%±0.0010 cd/m2)
10.00 to 1000 cd/m2 (±2%±0.0010 cd/m2)
|
Repeatability (2σ) *1
|
0.0050 to 0.0999 cd/m2 (1% + 0.0010 cd/m2)
0.1000 to 0.9999 cd/m2 (0.2% + 0.0010 cd/m2)
1.000 to 1000 cd/m2 (0.1%+0.0010 cd/m2)
|
0.0050 to 0.0999 cd/m2 (1% + 0.0010 cd/m2)
0.1000 to 0.9999 cd/m2 (0.2% + 0.0010 cd/m2)
1.000 to 1000 cd/m2 (0.1%+0.0010 cd/m2)
|
Chromaticity
|
Measurement range
|
0.0500 to 1000 cd/m²
|
0.0500 to 1000 cd/m²
|
Accuracy (white)*1
|
0.0500 to 4.999 cd/m2 (±0.005 for white)
5.000 to 19.99 cd/m2 (±0.004 for white)
20.00 to 1000 cd/m2 (±0.003 for white)
120 cd/m2 (±0.002 for white)
(±0.004 for monochrome)*2
|
0.0500 to 4.999 cd/m2 (±0.005 for white)
5.000 to 19.99 cd/m2 (±0.004 for white)
20.00 to 1000 cd/m2 (±0.003 for white)
120 cd/m2 (±0.002 for white)
(±0.004 for monochrome)*2
|
Repeatability (2σ) *1
|
0.0500 to 0.0999 cd/m²: 0.010
0.1000 to 0.1999 cd/m2: 0.004
0.2000 to 0.4999 cd/m²: 0.001
0.5000 to 1000 cd/m²: 0.001
|
0.0500 to 0.0999 cd/m²: 0.010
0.1000 to 0.1999 cd/m2: 0.004
0.2000 to 0.4999 cd/m²: 0.001
0.5000 to 1000 cd/m²: 0.001
|
Flicker Contrast method
|
Measurement range
|
-
|
5 cd/m2 or higher
|
Display range
|
-
|
0.0 ~ 999.9 %
|
Accuracy
|
-
|
±1 % (Flicker frequency: 30 Hz AC/DC 10% sine wave)
±2 % (Flicker frequency: 60 Hz AC/DC 10% sine wave)
|
Repeatability(2σ)
|
-
|
1 % (Flicker frequency: 20 to 65 Hz AC/DC 10% sine wave)
|
Flicker JEITA method *3
|
Measurement range
|
-
|
5 cd/m2 or higher
|
Accuracy
|
-
|
±0.5 dB (Flicker frequency: 30 Hz AC/DC 4% (-40 dB) sine wave)
±1.0 dB (Flicker frequency: 30 Hz AC/DC 1.2% (-50 dB) sine wave)
|
Repeatability(2σ)
|
-
|
0.1 dB (Flicker frequency: 30 Hz AC/DC 4% (-40 dB) sine wave)
0.3 dB (Flicker frequency: 30 Hz AC/DC 1.2% (-50 dB) sine wave)
|
Measurement speed (USB)
|
xyLv
|
0.0050 to 0.0999 cd/m²: 4 times/sec.
0.1000 to 1.999 cd/m²: 5 times/sec.
2.000 cd/m² to 1000 cd/m²: 20 times/sec.
|
0.0050 to 0.0999 cd/m²: 4 times/sec.
0.1000 to 1.999 cd/m²: 5 times/sec.
2.000 cd/m² to 1000 cd/m²: 20 times/sec.
|
Flicker (Contrast)
|
-
|
16(16) times/sec.
|
Flicker JEITA *3
|
-
|
0.5 (0.3)times/sec. *5
|
Display
|
Digital
|
xyLv, XYZ, TΔuvLv, u'v'Lv, RGB analyze
|
xyLV, TΔuvLV, RGB analyze, XYZ, u'v'LV, Flicker (Contrast method) *3
|
Analog
|
ΔxΔyΔLv, R/G B/G ΔG, ΔR B/R G/R
|
ΔxΔyΔLV, R/G B/G ΔG, ΔR B/R G/R, Flicker (Contrast method) *3
|
LCD
|
16 characters × 2 lines (with backlight)
|
SYNC mode
|
NTSC, PAL, EXT, UNIV, INT
|
Object Under Measurement
|
Vertical synchronous frequency: 40 to 200Hz
|
Vertical synchronization frequency: 40 to 200 Hz (Luminance or chromaticity measurement), 40 to 130 Hz (Flicker measurement)
|
Memory channels
|
100 channels
|
Analyzer function
|
Standard function
|
Interface
|
USB or RS-232c (38,400 bps maximum)
|
Multi-point measurement Max.
|
5 points (using optional 4-Probe Expansion Board CA-B15)
|
Operating temperature/humidity range
|
Temperature: 10 to 28°C; Relative humidity 70% or less with no condensation. Luminance change: ±2% ±1 digit of reading for white; Chromaticity change: ±0.002 for white, ±0.006 for monochrome from reading of Konica Minolta's standard LCD *1, 120.0 cd/m² at 23°C 40%RH
|
Storage temperature/humidity range
|
0 to 28°C: Relative humidity 70% or less with no condensation; 28 to 40°C: Relative humidity 40% or less with no condensation
|
Input voltage range
|
100 to 240V AC, 50/60Hz, 50VA
|
Dimensions (W × H × D)
|
Main body: 340 × 127 × 216mm
Probe: Ø49 × 208mm
|
Main body: 340 × 127 × 216mm
Probe: Ø49×208 mm
|
Weight
|
Main body: 3.58kg; Probe: 530g
|
Main body: 3.58kg; Probe: 530 g
|
*1 The chromaticity and luminance are measured under Konica Minolta's condition (standard LCD(6500K, 9300K) is used). *2 The luminance for monochrome is measured when the reading of luminance for white is 120 cd/m2 . *3 Measurement of flicker (JEITA method) is supported by SDK software. *4 Measuring probe connected to probe connector P1 only, used USB(used RS-232C Baud rate: 38400 bps) *5 Measured by Konica Minolta's PC (P3-600 MHz)
|