Fluke 5800A Oscilloscope Calibrator
]
Product Name:
Fluke 5800A Oscilloscope Calibrator
Product Model:
Fluke 5800A Oscilloscope Calibrator
Manufacturer:
Fluke
file:
No Related Documents
Product Description
The Fluke Model 5800A Oscilloscope Calibrator (Figure 1-1) is a precise instrument that
calibrates analog and digital oscilloscopes.
Fluke 5800A Oscilloscope Calibrator
|
5800A Mainframe Specifications
|
|
Voltage DC
|
|
Range:
|
0 to ±6.6V (50Ω)
|
|
Range:
|
±0.25%
|
|
|
------------
|
|
Range:
|
0 to ±130V (1 MΩ)
|
|
Best 1 Year Specification:
|
±0.025%
|
|
|
Voltage AC (Squarewave
|
|
Range:
|
: 0 to ±6.6V (50Ω)
|
|
Best 1 Year Specification:
|
±0.25%
|
|
|
--------
|
|
Range
|
±1 mV to ±130V p-p (1 MΩ)
|
|
Best 1 Year Specification:
|
±0.05%
|
|
|
Fast Edge
|
|
Range:
|
4 mV to 2.5V p-p (50Ω)
|
|
Best 1 Year Specification:
|
300 ps Rise Time
|
|
|
Leveled Sinewave
|
|
Range:
|
50 kHz to 600 MHz
|
|
Best 1 Year Specification:
|
50 kHz to 600 MHz
|
|
|
±1.5% Flatness
|
|
|
Time Markers
|
|
Range:
|
5s to 500 ps
|
|
|
Spike, square, 20% pulse, sine
|
|
Best 1 Year Specification:
|
±1 ppm
|
|
|
Wave Generator
|
|
Range:
|
Range: 1.8 mV to 55V p-p (1 MΩ)
|
|
|
1.8 mV to 2.5V p-p (50Ω)
|
|
|
0.01 Hz to 100 kHz
|
|
|
Square, Sine, Triangle
|
|
Best 1 Year Specification:
|
±3
|
|
|
Pulse Generator
|
|
Range:
|
15 mV to 1.5V amplitude
|
|
Best 1 Year Specification:
|
1 ns to 500 ns width ±5%
|
|
Best 1 Year Specification:
|
1 ns to 500 ns width ±5%
|
|
|
DC Voltage Measurement
|
|
Range:
|
0 to ±6.6V
|
|
Best 1 Year Specification:
|
: ±0.05%
|
|
|
Current DC
|
|
Range:
|
±100 mA to ±100 mA
|
|
Best 1 Year Specification:
|
± 0.25%
|
|
|
Current AC
|
|
Range:
|
100 mA p-p to 100 mA p-p
|
|
|
10 Hz to 100 kHz
|
|
|
Squarewave
|
|
Best 1 Year Specification:
|
±0.25%
|
|
|
TV Trigger
|
|
|
Input Impedance Measurement
|
|
Range:
|
40Ω to 1.5 MΩ
|
|
Best 1 Year Specification:
|
±0.1%
|
|
|
--------
|
|
Range:
|
5 pF to 50 pF
|
|
Best 1 Year Specification:
|
±5%
|
|
|
Overload Measurement
|
|
|
5V to 9V (dc or ac squarewave), 5 to 60 seconds
|
|
|
5800A with GHz Option
|
|
Voltage DC
|
|
Range:
|
0 to ±6.6V (50Ω)
|
|
Range:
|
±0.25%
|
|
|
------------
|
|
Range:
|
0 to ±130V (1 MΩ)
|
|
Best 1 Year Specification:
|
±0.025%
|
|
|
Voltage AC (Squarewave
|
|
Range:
|
: 0 to ±6.6V (50Ω)
|
|
Best 1 Year Specification:
|
±0.25%
|
|
|
--------
|
|
Range
|
±1 mV to ±130V p-p (1 MΩ)
|
|
Best 1 Year Specification:
|
±0.05%
|
|
|
Fast Edge
|
|
Range:
|
4 mV to 2.5V p-p (50Ω)
|
|
Best 1 Year Specification:
|
300 ps Rise Time
|
|
|
Leveled Sinewave
|
|
Range:
|
50 kHz to 2.1 GHz (with Option)
|
|
Best 1 Year Specification:
|
50 kHz to 2.1 GHz ± 1.5% Flatness
|
|
|
Time Markers
|
|
Range:
|
5s to 500 ps
|
|
|
Spike, square, 20% pulse, sine
|
|
Best 1 Year Specification:
|
±1 ppm
|
|
|
Wave Generator
|
|
Range:
|
Range: 1.8 mV to 55V p-p (1 MΩ)
|
|
|
1.8 mV to 2.5V p-p (50Ω)
|
|
|
0.01 Hz to 100 kHz
|
|
|
Square, Sine, Triangle
|
|
Best 1 Year Specification:
|
±3
|
|
|
Pulse Generator
|
|
Range:
|
15 mV to 1.5V amplitude
|
|
Best 1 Year Specification:
|
1 ns to 500 ns width ±5%
|
|
|
20 ms to 200 ns period ± 1 ppm
|
|
|
DC Voltage Measurement
|
|
Range:
|
0 to ±6.6V
|
|
Best 1 Year Specification:
|
: ±0.05%
|
|
|
Current DC
|
|
Range:
|
±100 mA to ±100 mA
|
|
Best 1 Year Specification:
|
± 0.25%
|
|
|
Current AC
|
|
Range:
|
100 mA p-p to 100 mA p-p
|
|
|
10 Hz to 100 kHz
|
|
|
Squarewave
|
|
Best 1 Year Specification:
|
±0.25%
|
|
|
TV Trigger
|
|
|
Input Impedance Measurement
|
|
Range:
|
40Ω to 1.5 MΩ
|
|
Best 1 Year Specification:
|
±0.1%
|
|
|
--------
|
|
Range:
|
5 pF to 50 pF
|
|
Best 1 Year Specification:
|
±5%
|
|
|
Overload Measurement
|
|
|
5V to 9V (dc or ac squarewave), 5 to 60 seconds
|
|
|
General Specifications
|
|
Warm up Time
|
|
|
Twice the time since last warmed up, to a maximum of 30 minutes
|
|
|
Settling Time
|
|
|
5 seconds or faster for all functions and ranges
|
|
|
Standard Interfaces
|
|
|
Temperature Performance
|
|
Operating:
|
0oC to 50oC
|
|
Calibration (tcal):
|
15oC to 35oC
|
|
Storage:
|
-20oC to 70oC
|
|
|
Electromagnetic Compatibility
|
|
|
Designed to operate in Standard Laboratory environments where the Electromagnetic environment is highly controlled. If used in areas with Electromagnetic fields > 1 V/m, there could be errors in output values.
|
|
|
Temperature Coefficient
|
|
|
Temperature Coefficient for temperatures outside tcal + 5 oC is 0.1X/oC of 1-year specifications.
|
|
|
Relative Humidity
|
|
Operating:
|
< 80% to 30 oC, < 70% to 40 oC, < 40% to 50 oC
|
|
|
Altitude
|
|
Non-operating:
|
12,200m (40,000 ft) maximum
|
|
Operating:
|
3,050m (10,000 ft) maximum
|
|
|
Safety
|
|
|
Designed to comply with IEC 1010-1 (1992-1); ANSI/ISA-S82.01-1994; CAN/CSA-C22.2 No. 1010, 1-92
|
|
|
Analog Low Isolation
|
|
|
EMC
|
|
|
Line Power
|
|
|
Line Voltage (selectable): 100V, 120V, 220V, 240V Line Frequency: 47 to 63 Hz Line Voltage Variation: + 10% about line voltage setting
|
|
|
Power Consumption
|
|
|
Dimensions
|
|
Height:
|
17.8 cm (7 in), standard rack increment, plus 1.5 cm (0.6 in) for feet on bottom of unit
|
|
Width:
|
43.2 cm (17 in), standard rack width
|
|
Depth:
|
47.3 cm (18.6 in) overall
|
|
Leave a message for product