Technical indicators
Frequency characteristics
Measurement frequency range :9kHz~ 2.6ghz
Resolution of center frequency setting :1Hz
Center frequency display accuracy:
The span is less than or equal to 2MHz
Plus or minus (2% of the span + center frequency * reference oscillator accuracy +50kHz) span >2MHz
Reference oscillator: by internal or external input switch (10 MHz)
Internal reference oscillator stability: plus or minus 2 x 10?Per day, plus or minus 2 times 10 minus 8 per year
Frequency span:
Linear mode: 1kHz~ 2.6ghz and 0
Logarithmic mode: choose the decimal system of 1,2 or 3 between 10kHz and 1000MHz
Frequency span accuracy:
It's less than or equal to the plus or minus 3% of the span, >, 2MHz
The span is greater than or equal to 2MHz
Frequency stability:
Residual FM:
50khzp-p or less span >10MHz
2khzp-p or less 10MHz is greater than or equal to >2MHz span
20kHzp minus p or less is less than or equal to 2MHz
Frequency drift :300Hz/min or less
The span of 2 MHZ or less
Sideband noise: offset at 20kHz less than or equal to 3.0ghz is less than or equal to -105dbc /Hz
At 20kHz, the deviation is less than or equal to 3.6ghz and less than or equal to minus 101dBc/Hz
Resolution:
3dB bandwidth: 30Hz to 1MHz, adjustable in 1 to 3 steps
6dB bandwidth: 200Hz,9kHz,120kHz
Selective: 15:1 or less (60 db: 3 db)
Bandwidth accuracy: less than 20%
Cursor accuracy:
Standard mode: center frequency display accuracy + span accuracy
Counter mode: display frequency x reference oscillator precision plus or minus 1 digit
Range of indicators
Range of amplitude measurement: -130dbm ~+25dBm