Main functions:
Heat treatment furnace temperature uniformity test, temperature and humidity test equipment temperature and humidity test, data automatic collection, automatic processing, automatic printing test.
System features:
The system is composed of laptop, multi-channel data collector or high-precision digital measuring instrument with multi-channel conversion function, which is easy to carry.
Automatic reference compensation and sensor calibration data correction.
Heat treatment furnace temperature uniformity test can be used S, R, B, J, K, N, E, T thermocouple.
The test method and requirements conform to GB/ t9452-2003 standard.
Temperature and humidity environment test equipment, temperature measurement PT100 thermal resistance or T thermocouple, humidity measurement with dry and wet ball or humidity sensor.
The test method and requirements conform to jjf-2003 test specification.
Technical indicators:
Thermocouple channel: 40
Thermal resistance channel: 20
Thermocouple measurement accuracy < 0.2 ℃
Thermal resistance measurement accuracy < 0.06 ℃ (PT100 platinum resistance)