Measurement Lines
|
Single-phase/two-wires to three-phase/four-wires (Through the use of various input units)
|
Measurement Items
|
(When using 9600 / 9601 / 9602, optional)
Voltage, current, voltage/current peak, effective/reactive/apparent power, power factor, phase, frequency, current/power integration, load rate, efficiency
(When using 9603, optional)
Voltage, torque, RPM, frequency, motor output
(When using 9605, optional)
Harmonic, waveform, voltage fluctuation / flicker measurement function
|
Measurement Ranges
|
Voltage: 6.0000 V to 1.0000 kV (depends on use of the input unit)
Current: 200.00 mA to 500.00 A (depends on use of the input unit)
Power: Depends on combination of voltage and current ranges
|
Integration Range
|
0 to ±999999 TAh/ TWh, (integration time up to 10000 hours)
|
Basic Accuracy
|
±0.1 % rdg. ±0.1 % f.s. (voltage, current, power, at 45 to 66Hz)
Note: When used together with the 9270 to 9272, and 9277 to 9279, resulting accuracy is the sum of that indicated above and the accuracy of these clamp-on sensors
|
Frequency Characteristics
|
Used with 9600: DC, 0.5 Hz to 1 MHz
Used with 9601: 5 Hz to 100 kHz
Used with 9602: DC, 0.5 Hz to 200 kHz (depends on the clamp-on sensor accuracy)
|
Signal Output
|
Analog level: Voltage, current, active power, 5 V DC f.s.
Waveform monitor: Voltage, current, 1 V rms f.s.
D/A output: Outputs 8 arbitrarily selected items, ±5 V DC f.s.
|
Other Functions
|
6.4 inch TFT color LCD, RMS/MEAN rectification, FDD, GP-IB/RS-232C interface, scaling, averaging
|
Power Supply
|
100 / 120 / 200 / 230 V AC (switched automatically), 50/60 Hz
|
Dimensions
|
430 mm(16.93 in)W × 150 mm(5.91 in)H × 370 mm(14.57 in)D
|
Weight
|
15 kg (529.1 oz) ...with all options
|
Accessories
|
Power cord (1), Connector (1)
|