Specifications:
Continuous Operation:
Continuous measurement of power, current, peak-to-peak voltage (closed load only), and crest factor
Single Operation:
Single measurement after the set delay time of the ESU output of power, current, peak-to-peak voltage (closed load only), and crest factor
Power Distribution:
Automatic measurement of power, current, peak-to-peak voltage (closed load only), and crest factor through a user-selectable load range
RF Leakage current:
Provides connections and load configurations to measure HF leakage from both grounded and isolated equipment
RECQM:
Test the return electrode control quality monitoring using the QA-ES internal loads
Manual remote:
Via Ansur test-automation software
Measurement:
True RMS value of applied waveform
RMS Bandwidth:
-
Instrumentation Only: 30 Hz to 10 MHz (-3 dB)
-
With loads: 30 Hz to 2.5 MHz (-3 dB)
Low Frequency Filter:
100 Hz filter to avoid low-frequency disturbance or interference
Current Accuracy:
20 mA to 2200 mA (5 % of reading + 2 mA) for loads from 100 to 2000 Ω
Load Resistance:
10 Ω to 2500 Ω in step of 25 Ω (@ dc), 2500 Ω to 5200 Ω in step of 100 Ω (@ dc)
Additional Fixed Load:
200 Ω 400 W for 30 s; max 15 % duty cycle
Crest Factor:
The higher of the two peak voltage measurements is used for computation. Range: 1.4 to 16 (V peak/V rms)
Foot-Switch Output:
The foot switch output can be used to trigger the electrosurgical unit
Peak-to-Peak Voltage:
0 kV to 10 kV (closed load only) accuracy: 10 % Note: Measurement is taken between the active and dispersive electrodes with closed load only.
Oscilloscope Output:
5 V/A uncalibrated, 100 mA RF current minimum input
Ansur QA-ES Plug-In Remote Control:
All functions and tests in QA-ES may be performed from the PC.
User-Programmable Test Sequences:
Allows unlimited numbers of test sequences with user-programmable templates and test limits. These tests include power distribution test, output test, HF leakage, and RECQM verification.
Storage and Recall:
Protocol formats and data may be stored, recalled, printed out, or transferred.
Temperature:
-
Operating: 15 ºC to 35 ºC (59 ºF to 95 ºF)
-
Storage: 0 ºC to 50 ºC (32 ºF to 122 ºF)
Display:
LCD graphic display
Alphanumeric Format:
8 lines x 40 characters
Graphic Mode:
240 x 64 pixel matrix
Display Control:
5 F-keys, enter, cancel, control knob, and an encoder
Data Input/Outputs:
Parallel printer port and bidirectional RS232
Power:
115/230 VAC; 48 to 66 Hz, 35 VA
Dimensions (WxDxH):
34.2 cm x 39.5 cm x 13.2 cm (13.5 in x 15.6 in x 5.2 in )
Weight: 9.8 kg (21.6 lb)
Key Features
-
| Automatic power distribution measurement
-
including power
-
current
-
peak-to-peak voltage closed load only)
-
and crest factorOscilloscope outputHigh-frequency leakage measurements with accuracy of ±5 % of reading128 internal user-selectable test loads from 10 \u03a9 to 5200 \u03a9Foot-switch output for triggering the ESU under testAnsur QA-ES software plug-in for automated test protocolsLarge displayRS-232 and Centronic-printer interface "